
GA 01.202/9.02 - 03/96
Description
1.2 Supplied Equipment
The equipment supplied with the TRIVAC-B pump inclu-
des:
Pump with motor, including initial filling of N 62 or
HE-200 oil, or Artic oil (Cat. Nos. 113 29 and 113 39),
or Anderol RCFE 96 N (Cat. Nos. 113 33).
1 centering ring,
1 centering ring with dirt trap,
2 clamping rings DN 25 KF.
As protection during shipment, the connection ports are
each blanked off by rubber diaphragms and supporting
rings.
5
TRIVAC-B pumps with single-phase AC motor are sup-
plied ready to operate with switch, built-in thermal motor
protection switch, mains cable (2 m) and mains plug.
For TRIVAC-B pumps with three-phase AC motor, the
switch, motor protection switch, mains cable etc. are not
included.
Fig. 3 Schematic of the lubricating system
Fig. 4 Hydropneumatic anti-suckback valve
Key to Fig. 3
1 Accessories
2 Bearings
3 Non-return valve
4 Pump chamber of the TRIVAC
5 Oil reservoir
6 Oil pump
Key to Fig. 4
1 Oil reservoir 5 Valve seat
2 Spring 6 Valve disk
3 Control piston 7 Gas inlet
4 Anti-suckback piston
Comentarios a estos manuales